Warning: session_start(): Cannot send session cookie - headers already sent by (output started at /home/content/29/14032429/html/system/database/mysql.php:6) in /home/content/29/14032429/html/system/library/session.php on line 12Warning: session_start(): Cannot send session cache limiter - headers already sent (output started at /home/content/29/14032429/html/system/database/mysql.php:6) in /home/content/29/14032429/html/system/library/session.php on line 12Warning: Cannot modify header information - headers already sent by (output started at /home/content/29/14032429/html/system/database/mysql.php:6) in /home/content/29/14032429/html/index.php on line 179Warning: Cannot modify header information - headers already sent by (output started at /home/content/29/14032429/html/system/database/mysql.php:6) in /home/content/29/14032429/html/system/library/currency.php on line 45 LaserGas II MP Wallmount
Specials

LaserGas II MP Wallmount

Product Code: NEO
Availability: In Stock
Qty: +
-
  - OR -  

LaserGas II MP Cabinet includes a heated multipass (Herriott) cell as an additional feature. This is used for applications where the gas sample must be heated to avoid condensation in the measurement cell.

Because the MP has no other requirements than power and sample gas tubes during installation, it is a turn-key instrument. To avoid failure in optical parts in the Multipass cell, cleanliness of the sample gas must be ensured. Filtering the sample gas in an appropriate extractive system may be required for some applications.

  • Short response time ( 2 – 10 sec)
  • Very low detection limits (ppb for most gases)
  • Heated cell up to 200 °C
  • No cross interference from background gases
  • Stable calibration, no zero drift
  • Offline gas analysis in controlled environment
  • No moving parts, no consumables
SINGLE GASES DUAL GASES
C2H2 H2O NH3 + H2O
C2H3CL H2S HCL+ H20
C2H4O HCl HF+H2O
C3H6 HCN CO+CO2
CH20 HF CO+H2O
CH2CL2 N2O CO+CH4
CH3I NH3 O2+TEMP
CH4 NO CO+TEMP
CO NO2 H2S + CO2
CO2 O2  

Specifications
Optical path length 0.6 – 12 m (1.9 – 39 Feet) (depends on cell version)

Response time typically 2 – 10 sec (depending on cell and sample gas flow) 

Averaging time Rolling average from 2 seconds to 24 hours (exp. decay) 

Repeatability +/- Detection limit or +/- 1% of reading, whichever is greater

Linearity < 1%

Environmental Conditions
Operating temperature -20 °C to +55°C (-4°F to 131°F)

Storage temperature -20 °C to +55 °C (-4°F to 131°F)

Protection classification IP65 (Cabinet), IP20 (19” Rack)

Inputs / Outputs
Analog output (3) 4 – 20 mA current loop

Digital output

RS - 232 format, Optional 10 or 10/100 Base T Ethernet, Optional fiber optic (ASCII - format)

Relay output (3) High gas-, Maintenance-, Warning - and Fault relays (normally closed-circuit relays)

Analog input 4 – 20 mA process temperature and pressure reading


Ratings
Input power supply unit 100 – 240 VAC, 50/60 Hz, 0.36 – 0.26 A

Output power supply unit 24 VDC, 900 – 1000 mA

Input transmitter unit 18 – 36 VDC, max. 20 W

4 – 20 mA output 500 Ohm max. Isolated

Relay output 1 A at 30 V DC/AC


Installation and Operation
Gas inlet / outlet 6 mm or 1/4 “ Swagelok (other dimensions on request)

Sample gas flow recommended 5 – 10 l/min

Sample inlet pressure 0.2 – 2.0 Bar abs (2.9 – 29 PSIA)

Purging of laser chamber (optional) Dry and oil-free pressurized air or gas, Nitrogen for O2 and H2O applications

Purge flow maximum 0.5 l/min


Maintenance
Visual inspection Recommended every 6 – 12 months (no consumables needed)

Remote instrument check by Ethernet connection or external modem possible

Calibration Check recommended every 12 months


Security
Laser class Class 1 according to IEC 60825-1

CE Certified, conformant with LVD 73/23/EEC, including 93/68/EEC EMC Conformant with directive 2004/108/EC

Explosion Protection (optional) Note: Only for Cabinet version
ATEX zone 2 II 3 G Ex nA nC op is Gc IIC T4, II 3 D Ex td A22 IP65 T100°C

CSA Class I, Div. 2, Groups A, B, C and D; Temp. Code T4; non-incendive

Dimension and Weight
Cabinet version 500 x 510 x 215 mm, 18.4 kg

19" Rack version 480 x 480 x 135 mm, 10 – 14 kg (dependent on measurement cell)